beautypg.com

Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 263

background image

Model 9110EH Instruction Manual APPENDIX A

Signal Name

Bit or Channel

Number

Description

HVPS_VOLTAGE

1

HV power supply output

PMT_TEMP 2

PMT

temperature

3

Spare

4

Temperature

MUX

5

Spare

6

Spare

SAMPLE_PRESSURE 7

Sample

pressure

RCELL_PRESSURE 8

Reaction

cell

pressure

REF_4096_MV

9

4.096V reference from MAX6241

OZONE_FLOW

10

Ozone flow rate

TEST_INPUT_11

11

Diagnostic test input

CONV_TEMP 12

Converter

temperature

TEST_INPUT_13

13

Diagnostic test input

14

DAC

loopback

MUX

REF_GND 15

Ground

reference

Rear board temperature MUX analog inputs, MUX default I/O address 326 hex

BOX_TEMP 0

Internal

box

temperature

RCELL_TEMP 1

Reaction

cell

temperature

IZS_TEMP 2

IZS

temperature

3

Spare

4

Spare

TEMP_INPUT_5

5

Diagnostic temperature input

TEMP_INPUT_6

6

Diagnostic temperature input

MANIFOLD_TEMP

2

7

Bypass or dilution manifold temperature

DAC_CHAN_0

0

DAC channel 0 loopback

1

DAC channel 1 loopback

DAC_CHAN_2

2

DAC channel 2 loopback

DAC_CHAN_3

3

DAC channel 3 loopback

Rear board analog outputs, default I/O address 327 hex

0

Concentration output #1 (NO

X

)

CONC_OUT_2

1

Concentration output #2 (NO)

CONC_OUT_3

2

Concentration output #3 (NO

2

)

3

Test measurement output

1

Hessen

protocol.

2

Factory

option.

Rear board DAC MUX analog inputs, MUX default I/O address 327 hex

DAC_CHAN_1

CONC_OUT_1

TEST_OUTPUT

M9110EH Rev 0

249