Model 9110EH Instruction Manual Operating Instructions

Table 0-16: COM1 and COM2 DB-9 Pin Assignments

Pin #

COM1 (RS-232)

COM2 (RS-232)

COM2 (RS-485)

1

Not used

Not used

Not used

2

TRANSMIT DATA*

RECEIVE DATA

DATA -

3

RECEIVE DATA*

TRANSMIT DATA

DATA +

4

Not used

Not used

Not used

5

Signal Ground

Signal Ground

Signal Ground

6

Not used

Not used

Not used

7

DATA SET READY*

DATA SET READY

Not used

8

REQUEST TO SEND*
(=DTE Ready)

REQUEST TO SEND

Not used

9

Not used

Not used

Not used

* Configurable for COM1 at rear panel using the DTE-DCE switch

6.9.5. DTE and DCE Communication

RS-232 was developed for allowing communications between data terminal equipment

(DTE) and data communication equipment (DCE). Basic terminals always fall into the DTE
category whereas modems are always considered DCE devices. The difference between the

two is the pin assignment of the Data Receive and Data Transmit functions. DTE devices
receive data on pin 2 and transmit data on pin 3, DCE devices receive data on pin 3 and

transmit data on pin 2.

To allow the analyzer to be used with terminals (DTE), modems (DCE) and computers
(which can be either), a switch mounted below the serial ports on the rear panel allows the

user to switch between the two functions (for COM1 only).

6.9.6. COM Port Communication Modes

Each of the analyzer’s serial ports can be configured to operate in a number of different
modes, which are listed in Table 0-17 and which can be combined by adding the Mode ID

numbers. For example, quiet mode, computer mode and internet-enabled mode would carry
a combined mode ID of 11, the standard configuration for Ethernet functionality on the
M9110EH

COM2 port. Note that each COM port needs to be configured independently.

M9110EH Rev 0