Connect zero air and span gases to the analyzer, Set expected no and nox span gas concentrations, Set expected no and no – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual
Page 131: Span gas concentrations, Figure 7-2, Set expected, Pan gas concentrations, No and no, Model 9110eh
Model 9110EH Instruction Manual Calibration Procedures
7.2.1. Connect Zero Air and Span Gases to the Analyzer
Pump
MODEL
9110EH
Sample
Zero Air
Exhaust
Span
MODEL 700 Gas
Dilution
Calibrator
(with O
3
generator
option)
MODEL 701
Zero Air
Generator
Vent excess gas
Calibrated NO
gas
(high
concentration)
Source of
sample gas
removed
during
calibration
Valve
OR
Needle valve to
control flow
MODEL
9110EH
Sample
Zero Air
Exhaust
Span
VENT
Calibrated NO
gas
(At span gas
concentration)
Source of
SAMPLE Gas
removed
during
calibration
Pump
MODEL 701
Zero Air
Generator
Figure 7-2: Setup for Manual Calibration without Z/S valve or IZS Option
7.2.2. Set Expected
NO and NO
X
S
pan Gas Concentrations
SAMPLE RANGE = 500.0 PPB NOX= X.X
< TST TST > CAL
SETUP
This sequence causes the
analyzer to prompt for the
expected NO
x
span
concentration.
The NO
X
& NO span concentration
values automatically default to
400.0 Conc.
To change this value to the
concentration of the span gas, enter
the number by pressing the key
under each digit until the value is
set.
EXIT ignores the new setting and
returns to the
CONCENTRATION MENU.
ENTR accepts the new setting and
returns to the
CONCENTRATION MENU..
M-P CAL CONCENTRATION MENU
NOX NO CONV
EXIT
M-P CAL NOX SPAN CONC: 400.0 Conc
0 0 4 0 0 .0 ENTR EXIT
M-P CAL CONCENTRATION MENU
NOX NO CONV
EXIT
M-P CAL NO SPAN CONC: 450.0 Conc
0 0 4 5 0 .0 ENTR EXIT
M-P CAL RANGE = 500.0 PPB NO = X.X
< TST TST > ZERO CONC
EXIT
M9110EH Rev 0
117