beautypg.com

Figure 7-1, Gas supply setup for determination of no, Conversion efficiency – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 129: Model 9110eh

background image

Model 9110EH Instruction Manual Calibration Procedures

Pump

MODEL

9110EH

Sample

Zero Air

Exhaust

Span

Gas Dilution

Calibrator

(with O

3

generator

option)


Zero Air Source

VENT

Calibrated

NO

2

gas

(high

concentration)

Source of

SAMPLE Gas

removed

during

procedure

Figure 7-1: Gas Supply Setup for Determination of NO

2

Conversion Efficiency

Step two is to set the expected NO

2

span gas concentration:

M-P CAL CONCENTRATION MENU

NOX NO CONV

EXIT

SAMPLE A RANGE TO CAL: LOW NOX=X.X

LOW HIGH

ENTR

SAMPLE RANGE = 500.0 PPB NOX=X.X

< TST TST > ZERO CONC

EXIT

M-P CAL

CO SPAN CONC: 400.0 Conc

0 0 4 0 0 .0 ENTR EXIT

The expected NO

2

span concentration value

defaults to 400.0 Conc.

Make sure that you specify the actual

concentration value of the span gas. To

change this value to the concentration of the

span gas, enter the number by pressing the

key under each digit until the value is visble.

M-P CAL CONVERTER EFFICIENCY MENU

NO

2

CAL SET

EXIT

ENTR accepts the expected NO

2

span

concentration value. EXIT ignores the

expected NO

2

span concentration value.

SAMPLE RANGE = 500.0 PPB NOX=X.X

< TST TST > CAL

SETUP

This step only appears if the

analyzer’s reporting range is set

for AUTO range mode.

Select LOW and press ENTR.

Repeat entire procedure for

HIGH range.

M9110EH Rev 0

115