beautypg.com

Pneumatic connections, Table 3-1, Inlet / outlet connector nomenclature – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 27

background image

Model 9110EH Instruction Manual Getting Started

3.3. Pneumatic Connections

Sample and calibration gases should only come into contact with PTFE (Teflon), FEP, glass

or stainless steel materials. Figure 3-5 and Figure 3-6 illustrate the most common
configurations for gas supply and exhaust lines to the Model 9110EH Analyzer. Appendix D

contains more detailed pneumatic flow diagrams for the analyzer and its various
(pneumatically related) options.

CAUTION

To prevent dust from getting into the analyzer, it was shipped with small plugs

inserted into each of the pneumatic fittings on the rear panel. Make sure that all

dust plugs are removed before attaching exhaust and supply gas lines.

Please refer to Figure 3-3 for pneumatic connections at the rear panel and Table 3-1 for
nomenclature.

Table 3-1: Inlet / Outlet Connector Nomenclature

Rear Panel Label

Function

Sample

Connects the sample gas to the analyzer. When operating the analyzer

without zero span option, this is also the inlet for any calibration gases.

Exhaust

Connects the exhaust of the analyzer with the external vacuum pump.

Span

On Units with zero/span valve or IZS option installed, this port

connects the external calibration gas to the analyzer.

Zero Air

On Units with zero/span valve or IZS option installed, this port
connects the zero air gas or the zero air cartridge to the analyzer.

• Attach a sample inlet line to the sample inlet port. Ideally, the pressure of the

sample gas should be equal to ambient atmospheric pressure.

• In applications where the sample gas is received from a pressurized manifold, a vent

must be provided to equalize the sample gas with ambient atmospheric pressure
before it enters the analyzer. The vented gas needs to be routed outside the imme-

diate area or shelter surrounding the instrument.

CAUTION

Maximum pressure of any gas at the sample inlet should not exceed 1.5 in-Hg

above ambient pressure and ideally should equal ambient atmospheric pressure.

CAUTION

The exhaust from the external pump needs to be vented outside the

immediate area or shelter surrounding the instrument using a

maximum of 10 meters of 1/4” PTFE tubing.

M9110EH

Rev 0 13