Appendix a-5 – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual
Page 264
APPENDIX A Model 9110EH Instruction Manual
APPENDIX A-5: M9110EH Trigger Events and iDAS Functions, Revision C.8
Table A-5: M9110EH DAS Trigger Events, Revision C.8
Trigger Name
Description
Events that can be accessed through the front panel
ATIMER
Automatic timer expired
EXITZR
Exit zero calibration mode
EXITHS
Exit high span calibration mode
EXITMP
Exit multi-point calibration mode
SLPCHG
Slope and offset recalculated
Events that can be accessed through the remote interface
EXITDG
Exit diagnostic mode
AZEROW Auto-zero
warning
OFLOWW
Ozone flow warning
RPRESW
Reaction cell pressure warning
RTEMPW
Reaction cell temperature warning
MFTMPW
1
Bypass or dilution manifold temperature warning
IZTMPW
1
IZS temperature warning
CTEMPW
Converter temperature warning
PTEMPW
PMT temperature warning
SFLOWW Sample
flow
warning
BTEMPW
Box temperature warning
HVPSW
HV power supply warning
Table A-6: M9110EH iDAS Functions, Revision C.8
Name Description
Units
PMTDET
PMT detector reading
mV
NXSLP1 NO
X
slope for range #1
—
NXSLP2 NO
X
slope for range #2
—
NOSLP1
NO slope for range #1
—
NOSLP2
NO slope for range #2
—
NXOFS1 NO
X
offset for range #1
mV
NXOFS2 NO
X
offset for range #2
mV
NOOFS1
NO offset for range #1
mV
NOOFS2
NO offset for range #2
mV
NXZSC1 NO
X
concentration for range #1 during zero/span calibra-
tion, just before computing new slope and offset
ppb
2
NXZSC2 NO
X
concentration for range #2 during zero/span calibra-
tion, just before computing new slope and offset
ppb
2
NOZSC1
NO concentration for range #1 during zero/span calibration,
just before computing new slope and offset
ppb
2
250 M9110EH Rev 0