Model 9110EH Instruction Manual APPENDIX A

Setup Variable

Numeric

Units

Default

Value

Value

Range

Description

MODEM_INIT

“AT Y0 &D0
&H0 &I0

S0=2 &B0

&N6 &M0 E0
Q1 &W0”

Any
character

in the

allowed
character

set. Up

to 100
character

s long.

RS-232 COM1 modem initialization
string. Sent verbatim plus carriage

return to modem on power up or

manually. Enclose value in double
quotes (") when setting from the

RS-232 interface.

RS232_MODE2

BitFlag

11

0-65535

RS-232 COM2 mode flags.
(Same settings as RS232_MODE)

BAUD_RATE2 —

115200

300,

1200,
2400,

4800,

9600,
19200,

38400,

57600,

115200

RS-232 COM2 baud rate. Enclose

value in double quotes (") when
setting from the RS-232 interface.

MODEM_INIT2

“AT Y0 &D0

&H0 &I0
S0=2 &B0

&N6 &M0 E0

Q1 &W0”

Any

character
in the

allowed

character

set. Up
to 100

character

s long.

RS-232 COM2 modem initialization

string. Sent verbatim plus carriage
return to modem on power up or

manually. Enclose value in double

quotes (") when setting from the

RS-232 interface.

RS232_PASS Password

940331

0-
999999

RS-232 log on password.

MACHINE_ID

ID

200

0-9999

Unique ID number for instrument.

GATEWAY_IP_ADDR

1

“000.000.

000.000”

Any

string of
digits

and

decimal
points.

Gateway IP address.

INST_IP_ADDRESS

1

“000.000.

000.000”

Any

string of
digits

and

decimal
points.

Instrument IP address.

INST_SUBNET_MASK

1

— “255.255.

255.0”

Any

string of
digits

and

decimal

points.

Instrument subnet mask.

INST_TCP_PORT

1

3000

1024-

65535

Instrument TCP port.

M9110EH Rev 0

241