Figure 3-8, Status output connector, Table 3-2 – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual
Page 32: Analog output pin-outs, Table 3-3, Status output signals
Getting Started Model 9110EH Instruction Manual
Table 3-2: Analog Output Pin-Outs
Pin
Analog Output
Data Type
Voltage Signal
Current Signal
1
V Out
I Out +
2
A1 NO
X
Ground
I Out -
3
V Out
I Out +
4
A2 NO
Ground
I Out -
5
V Out
I Out +
6
A3 NO
2
Ground
I Out -
7
V Out
not available
8
A4 Test
Channel
Ground not
available
If you wish to utilize the analyzer’s status outputs to interface with a device that accepts
logic-level digital inputs, such as programmable logic controller (PLC) chips, you can access
them through a 12 pin connector on the analyzer’s rear panel labeled STATUS.
E
M
ITT
E
R BU
S
FOR P
IN
S
1
-8
STATUS
1 2 3 4 5 6 7 8 D
+
SY
ST
E
M
O
K
HIGH RAN
G
E
CO
N
C
V
A
LI
D
ZE
RO
CAL
S
P
A
N
CAL
DIA
G
NO
S
T
IC
MODE
LO
W
S
P
AN
Figure 3-8: Status Output Connector
Table 3-3: Status Output Signals
Pin #
Status
Condition (ON = Conducting)
1
SYSTEM OK
ON if no faults are present.
2
CONC VALID
ON if concentration measurement (NO, NO2 or NOx) is valid.
OFF any time the hold-off feature is active.
3
HIGH RANGE
ON if unit is in high range of the Auto Range Mode.
4
ZERO CAL
ON whenever the instrument is in ZERO point calibration mode.
5
SPAN CAL
ON whenever the instrument is in SPAN point calibration mode.
6
DIAG MODE
ON whenever the instrument is in diagnostic mode.
7 Unused
8 Unused
D
EMITTER BUS The emitters of the transistors on pins 1-8 are tied together.
Unused
+
DC POWER
+ 5 VDC, 300 mA (combined rating with Control Output, if used).
Digital
Ground
The ground level from the analyzer’s internal DC power supplies