Getting Started Model 9110EH Instruction Manual

Table 3-2: Analog Output Pin-Outs

Pin

Analog Output

Data Type

Voltage Signal

Current Signal

1

V Out

I Out +

2

A1 NO

X

Ground

I Out -

3

V Out

I Out +

4

A2 NO

Ground

I Out -

5

V Out

I Out +

6

A3 NO

2

Ground

I Out -

7

V Out

not available

8

A4 Test

Channel

Ground not

available

If you wish to utilize the analyzer’s status outputs to interface with a device that accepts

logic-level digital inputs, such as programmable logic controller (PLC) chips, you can access
them through a 12 pin connector on the analyzer’s rear panel labeled STATUS.

E

M

ITT

E

R BU

S

FOR P

IN

S

1

-8

STATUS

1 2 3 4 5 6 7 8 D

+

SY

ST

E

M

O

K

HIGH RAN

G

E

CO

N

C

V

A

LI

D

ZE

RO

CAL

S

P

A

N

CAL

DIA

G

NO

S

T

IC

MODE

LO

W

S

P

AN

Figure 3-8: Status Output Connector

Table 3-3: Status Output Signals

Pin #

Status

Condition (ON = Conducting)

1

SYSTEM OK

ON if no faults are present.

2

CONC VALID

ON if concentration measurement (NO, NO2 or NOx) is valid.
OFF any time the hold-off feature is active.

3

HIGH RANGE

ON if unit is in high range of the Auto Range Mode.

4

ZERO CAL

ON whenever the instrument is in ZERO point calibration mode.

5

SPAN CAL

ON whenever the instrument is in SPAN point calibration mode.

6

DIAG MODE

ON whenever the instrument is in diagnostic mode.

7 Unused

8 Unused

D

EMITTER BUS The emitters of the transistors on pins 1-8 are tied together.

Unused

+

DC POWER

+ 5 VDC, 300 mA (combined rating with Control Output, if used).

Digital
Ground

The ground level from the analyzer’s internal DC power supplies