Model 9110EH Instruction Manual Theory of Operation

Figure 10-18:

Keyboard and Display Interface Block Diagram

10.4. Software Operation

The M9110EH NO

X

analyzer’s core module is a high performance, 386-based microcomputer

running a version of DOS. On top of the DOS shell, special software developed by TAI
interprets user commands from various interfaces, performs procedures and tasks, stores

data in the CPU’s memory devices and calculates the concentrations of NO

X

in the sample

gas. Figure 10-19 shows a block diagram of this software functionality.

10.4.1. Adaptive Filter

The M9110EH NO

X

analyzer software processes sample gas concentration data through a built-

in adaptive filter. Unlike other analyzers that average the output signal over a fixed time

period, the M9110EH averages over a defined number of samples, with samples being about 8
seconds apart (reflecting the switching time of 4 s each for NO and NO

X

). This technique is

known as boxcar filtering. During operation, the software may automatically switch between
two different filters lengths based on the conditions at hand.

During constant or nearly constant concentrations, the software, by default, computes an
average of the last 42 samples, or approximately 5.6 minutes. This provides smooth and

stable readings and averages out a considerable amount of random noise for an overall less
noisy concentration reading.

If the filter detects rapid changes in concentration the filter reduces the averaging to only 6

samples or about 48 seconds to allow the analyzer to respond more quickly. Two conditions