Figure 7-3, Table 7-1, Izs option valve states with cal_on_no – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual
Page 135: Turned on, 9110eh
Model 9110EH Instruction Manual Calibration Procedures
permeation tube and through the NO
2
converter (Figure 3-11). It also programs the
analyzer to use the NO output from the NO
2
converter to calibrate the span value of both
NO and NO
X
.
NOTE
The IZS with NO
2
permeation tube has a limited accuracy of about ±5%. Whereas
this may be sufficient for informal calibration checks, this method of calibration is
not approved by the US EPA as a calibration source.
This feature should only be enabled when a span calibration or calibration check is
performed. While CAL-ON-NO2 is enabled, the NO
2
concentration will always be reported
as zero. This is because the gas is continuously routed through the NO
2
converter and the
analyzer’s firmware simulates calibration with NO gas. Since the instrument sees the same
concentration of NO during both NO and NO
X
cycles, it reports an NO
2
concentration of zero.
Table 7-1: IZS Option Valve States with CAL_ON_NO
2
Turned ON
Valve
Condition
Valve Port Connections
Sample/Cal
Open to zero/span valve
1 Æ 2
Zero/Span
Open to SPAN GAS inlet
1 Æ 2
NO/NOx Valve
Open to NO2 converter
1 Æ 2
AutoZero Valve
Cycles normally
N/A
Performing a manual calibration of the analyzer with an IZS option installed is similar to the
method described in Section 7.4, except that both zero air and span gas are supplied to the
analyzer from ambient air through the zero gas inlet.
1.
Connect the sources of zero air and span gas as shown in Figure 7-3.
Figure 7-3: Pneumatic Connections for Manual Calibration Checks with IZS
Set the expected NO and NO
X
values according to Section 7.2.
Perform the zero point calibration:
M9110EH Rev 0
121
9110EH