Appendix c, Repair questionnaire - m200e, Appendix c: repair questionnaire - model 9110eh – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual
Page 276
APPENDIX C Model 9110EH Instruction Manual
APPENDIX C: Repair Questionnaire -
Model 9110EH
Company: __________________Contact Name: _________ _______________________
Phone Number: _____________Fax Number: ___________ Email: __________________
Site Address: _______________ _____________________ _______________________
Can we connect to the instrument? If so, provide IP address or modem: _______________
Model Serial Number: ____ _____________________ Firmware revision: ________
The serial number can be found on the back of the instrument, the firmware revision is displayed in the upper left corner of the
display when pressing SETUP on the front panel (Example: C.3).
1. List all front panel error/warning messages: ___________ _______________________
__________________________ _____________________ _______________________
2. Please complete the following table: (Depending on options installed, not all test
parameters shown below may be available in your instrument)
Parameter
Recorded Value
Acceptable Value
Parameter
Recorded
Value
Acceptable Value
RANGE
ppb/ppm
50 ppb - 20 ppm
PMT TEMP
°C
7 ± 2
NO
X
STB
ppb/ppm ≤ 1 ppb with zero air
IZS TEMP
°C
50 ± 1
Sample FL
cm³/min
500 ± 50
MOLY TEMP
°C
315 ± 5
Ozone FL
cm³/min
80 ± 15
RCEL pressure
in-Hg-A
<10
PMT signal with
zero air
mV
-20 to 150
SAMP pressure
in-Hg-A
Ambient - ~1
PMT signal at
span gas conc
mV
ppb/ppm
0-5000
0-20000 ppb
NO SLOPE
1.0 ± 0.3
NORM PMT at
span gas conc
mV
ppb/ppm
0-5000
0-20000 ppb
NO OFFS
mV
-50 to 150
AZERO
mV
-20 to 150
NO
X
SLOPE
1.0 ± 0.3
HVPS
V
400-900
NOX OFF
mV
-50 to 150
RCELL TEMP
°C
50 ± 1
ETEST PMT
mV
2000 ± 1000
BOX TEMP
°C
PMT mV
Ambient + ~5
OTEST
2000 ± 1000
3. Has the analyzer been checked for leaks? Yes No ; for proper flows? Yes No
4. What are the failure symptoms? ____________________ _______________________
_____________________________ ________________________ __ Continue on back if necessary
5. Which tests have you done trying to solve the problem? _ _______________________
__________________________ _____________________ __
Continue on back if necessary
6. If possible, fax a portion of a strip chart or email a data file to customer service:
262 M9110EH Rev 0
FAX: 626.961.2538