beautypg.com

Appendix c, Repair questionnaire - m200e, Appendix c: repair questionnaire - model 9110eh – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 276

background image

APPENDIX C Model 9110EH Instruction Manual

APPENDIX C: Repair Questionnaire -

Model 9110EH

Company: __________________Contact Name: _________ _______________________
Phone Number: _____________Fax Number: ___________ Email: __________________
Site Address: _______________ _____________________ _______________________
Can we connect to the instrument? If so, provide IP address or modem: _______________
Model Serial Number: ____ _____________________ Firmware revision: ________

The serial number can be found on the back of the instrument, the firmware revision is displayed in the upper left corner of the

display when pressing SETUP on the front panel (Example: C.3).

1. List all front panel error/warning messages: ___________ _______________________

__________________________ _____________________ _______________________

2. Please complete the following table: (Depending on options installed, not all test
parameters shown below may be available in your instrument)

Parameter

Recorded Value

Acceptable Value

Parameter

Recorded

Value

Acceptable Value

RANGE

ppb/ppm

50 ppb - 20 ppm

PMT TEMP

°C

7 ± 2

NO

X

STB

ppb/ppm ≤ 1 ppb with zero air

IZS TEMP

°C

50 ± 1

Sample FL

cm³/min

500 ± 50

MOLY TEMP

°C

315 ± 5

Ozone FL

cm³/min

80 ± 15

RCEL pressure

in-Hg-A

<10

PMT signal with
zero air

mV

-20 to 150

SAMP pressure

in-Hg-A

Ambient - ~1

PMT signal at
span gas conc

mV

ppb/ppm

0-5000

0-20000 ppb

NO SLOPE

1.0 ± 0.3

NORM PMT at
span gas conc

mV

ppb/ppm

0-5000

0-20000 ppb

NO OFFS

mV

-50 to 150

AZERO

mV

-20 to 150

NO

X

SLOPE

1.0 ± 0.3

HVPS

V

400-900

NOX OFF

mV

-50 to 150

RCELL TEMP

°C

50 ± 1

ETEST PMT

mV

2000 ± 1000

BOX TEMP

°C

PMT mV

Ambient + ~5

OTEST

2000 ± 1000

3. Has the analyzer been checked for leaks? Yes No ; for proper flows? Yes No

4. What are the failure symptoms? ____________________ _______________________

_____________________________ ________________________ __ Continue on back if necessary

5. Which tests have you done trying to solve the problem? _ _______________________

__________________________ _____________________ __

Continue on back if necessary

6. If possible, fax a portion of a strip chart or email a data file to customer service:


262 M9110EH Rev 0

FAX: 626.961.2538