Manual analog output calibration, Figure 6-2, Setup for calibrating analog outputs – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual
Page 82: Table 6-11, Voltage tolerances for analog output calibration
Operating Instructions Model 9110EH Instruction Manual
6.7.3.3. Manual Analog Output Calibration
For highest accuracy, the voltages of the analog outputs can be manually calibrated. Note
that outputs configured for 0.1V full scale should always be calibrated manually. Calibration
is done through the instrument software with a voltmeter connected across the output
terminals (Figure 0-3). Adjustments are made using the front panel keys by setting the
zero-point first and then the span-point (Table 0-11). The software allows this adjustment
to be made in 100, 10 or 1 count increments.
Table 0-11: Voltage Tolerances for Analog Output Calibration
Full Scale
Zero Tolerance
Span Voltage
Span Tolerance
0.1 VDC
±0.0005V
90 mV
±0.001V
1 VDC
±0.001V
900 mV
±0.001V
5 VDC
±0.002V
4500 mV
±0.003V
10 VDC
±0.004V
4500 mV
±0.006V
V
+DC Gnd
V OUT +
V OUT -
V IN +
V IN -
Recording
Device
ANALYZER
See Table 3-1 for
pin assignments
of Analog Out
connector on the
rear panel
Figure 0-2: Setup for Calibrating Analog Outputs
To make these adjustments, the AOUT auto-calibration feature must be turned off
(Section 6.7.3). Then press SETUP-MORE-DIAG-ENTR-NEXT-NEXT and the following keys:
68 M9110EH Rev 0