Precision check, Precision check procedure, Certification of working standards – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 156

background image

EPA Protocol Calibration Model 9110EH Instruction Manual

• Automatic zero/span checks can be performed every night. See Section 7.8 of this

manual for setup and operation procedures.

• Zero/Span checks through remote contact closure can be initiated through remote

contact closures on the rear panel. See Section 7.7.3 of this manual.

• Zero/span checks can also be controlled through the RS-232 port. See Section 6.9 of

this manual for more details on setting up and using the analyzer’s RS-232 port.

8.7.2. Precision Check

A periodic check is used to assess the data for precision. A one-point precision check must

be carried out at least once every 2 weeks on each analyzer at an NO

2

concentration

between 0.08 and 0.10 ppm. The analyzer must be operated in its normal sampling mode,

and the precision test gas must pass through all filters, scrubbers, conditioners, and other
components used during normal ambient sampling. The standards from which precision

check test concentrations are obtained must be traceable to NIST-SRM. Those standards
used for calibration or auditing may be used.

8.7.3. Precision Check Procedure

Connect the analyzer to a precision gas that has an NO

2

concentration between 0.08 and

0.10 ppm. An NO

2

precision gas may be generated by either GPT or a NO

2

permeation tube.

If a precision check is made in conjunction with a zero/span check, it must be made prior to
any zero or span adjustments.

• Allow the analyzer to sample the precision gas until a stable trace is obtained.
• Record this value. NO and NO

X

precision checks should also be made if those data

are being reported. Information from the check procedure is used to assess the
precision of the monitoring data; see in Section 12 of the Q.A. Handbook

6

for

procedures for calculating and reporting precision.

8.8. Certification of Working Standards

The NO content of the NO working standard must be periodically assayed against NIST-
traceable NO or NO

2

standards. Any NO

2

impurity in the cylinder must also be assayed.

Certification of the NO working standard should be made on a quarterly basis or more
frequently, as required. Procedures are outlined below for certification against NO traceable

standard which is the simplest and most straightforward procedure.

To assure data of desired quality, two considerations are essential:

• The measurement process must be in statistical control at the time of the

measurement and;

• Any systematic errors, when combined with the random variation in the measure-

ment process, must result in a suitably small uncertainty.

Evidence of good quality data includes documentation of the quality control checks and the

independent audits of the measurement process by recording data on specific forms or on a
quality control chart and by using materials, instruments, and measurement procedures

that can be traced to appropriate standards of reference.

142 M9110EH Rev 0