EPA Protocol Calibration Model 9110EH Instruction Manual

F

]

[NO

x

F

=

]

[NO

T

STD

NO

OUT

Eq 8-8

Enter the respective concentrations using the procedure in Section 7.2. The expected span
concentrations need not be re-entered each time a calibration is performed unless they are

changed.

• Enter the expected NO

X

and NO span gas concentrations:

• Sample the generated concentration until the NO and the NO

X

responses have

stabilized.

• Span the instrument by the following the same method as Sections 7.2, 7.4 or 7.5:

The analog voltage output should measure 80% of the voltage range selected. (e.g.
4.00 VDC if 0-5V output is selected.) The readings on the front panel display should be

equal to the expected NO and NO

X

concentrations entered in the procedure given in

Sections 7.2, 7.4 or 7.5 above. See the Troubleshooting Chapter 11 if there are problems.

Also see the Calibration Quality Check procedure Section 7.9.

After the zero and the 80% URL points have been set, generate five approximately evenly
spaced calibration points between zero and 80% URL without further adjustment to the

instrument. Allow the instrument to sample these intermediate concentrations for about 10
minutes each and record the instrument NO and NO

X

responses.

Plot the analyzer NO and NO

X

responses versus the corresponding calculated concentrations

to obtain the calibration relationships. Determine the straight line of best fit (y = mx + b)

determined by the method of least squares.

After the best-fit line has been drawn for the NO and the NO

X

calibrations, determine

whether the analyzer response is linear. To be considered linear, no calibration point should
differ from the best-fit line by more than 2% of full scale.

8.3.3. GPT NO

2

Calibration

The M9110EH computes the NO

2

concentration by subtracting the NO from the NO

X

concentration. Unlike analog instruments, this difference is calculated by the M9110EH's

internal computer software. It is extremely unlikely that the NO

2

concentration will be in

error. Therefore this procedure is a confirmation that the NO

2

subtraction algorithm in the

computer is operating correctly.

NOTE

Do not make any adjustments to the instrument during this procedure.

Generate an NO concentration near 90% of the URL. Dilution air and O

3

generator air flows

should be the same as used in the calculation of specified conditions of the dynamic
parameter according to Section 8.2. Sample this NO concentration until the NO and NO

X

responses stabilize. Record the NO and NO

x

concentrations.

Second, turn on and adjust the O

3

generator in the calibrator to produce sufficient O

3

to

decrease the NO concentration to about 10% of full scale. This will be equivalent to 80% of
the URL of the NO

2

channel. After the analyzer responses stabilize, record the resultant NO,

NO

X

, and NO

2

concentrations.

138 M9110EH Rev 0