beautypg.com

Specifications, approvals and warranty, M200e operating specifications, M9110eh operating specifications – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 19: Table 2-1, Model 9110eh basic unit specifications

background image

Model 9110EH Instruction Manual M9110EH Documentation

2. SPECIFICATIONS, APPROVALS AND

WARRANTY

2.1. M9110EH Operating Specifications

Table 2-1: Model 9110EH Basic Unit Specifications

Min/Max Range
(Physical Analog Output)

Min: 0-50 ppb
Max: 0-20 ppm

Measurement Units

ppb, ppm, µg/m

3

, mg/m

3

(user selectable)

Zero Noise

1

0.2 ppb (RMS)

Span Noise

1

<0.5% of reading above 50 ppb

or 0.2 ppm, whichever is greater

Lower Detectable Limit

2

0.4

ppb

Zero Drift (24 hours)

<0.5 ppb (at constant temperature and voltage.)

Zero Drift (7 days)

1 ppb (at constant temperature and voltage.)

Span Drift (7 Days)

<0.5% of full Scale (at constant temperature and voltage.)

Linearity

1% of full scale

Precision

0.5% of reading

Temperature Coefficient

< 0.1% per °C

Voltage Coefficient

< 0.1% per V

Lag Time

1

20 s

Rise/Fall Time

1

95% in <60 s

Sample Flow Rate

500 cm

3

/min. ± 10%

Temperature Range

5 - 40 °C operating and EPA equivalency

Humidity Range

0-95% RH non-condensing

Dimensions H x W x D

18 cm x 43 cm x 61 cm (7" x 17" x 23.6")

Weight, Analyzer

18 kg (40 lbs)

Weight, Ext Pump Pack

7 kg (16 lbs)

AC Power Rating

100 V, 50/60 Hz (3.25A);
115 V, 60 Hz (3.0 A);
220 - 240 V, 50/60 Hz (2.5 A)

Power, Ext Pump

100 V, 50/60 Hz (3.25A); 115 V, 60 Hz (3.0 A);
220 - 240 V, 50/60 Hz (2.5 A)

Environmental

Installation category (over-voltage category) II; Pollution degree 2

Analog Outputs

4 outputs

Analog Output Ranges

All Outputs: 0.1 V, 1 V, 5 V or 10 V (user selectable)

Three outputs convertible to 4-20 mA isolated current loop.
All Ranges with 5% under/over range

Analog Output Resolution

1 part in 4096 of selected full-scale voltage (12 bit)

Status Outputs

8 Status outputs from opto-isolators, 7 defined, 1 spare

Control Inputs

6 Control inputs, 4 defined, 2 spare

Serial I/O

1 RS-232; 1 RS-485 or RS-232 (configurable)
Communication speed: 300 - 115200 baud (user selectable)

Certifications

USEPA: Reference Method Number RFNA 1194-099

CE: EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B Section
15.107 Class A, ICES-003 Class A (ANSI C63.4 1992) & AS/NZS 3548
(w/A1 & A2; 97) Class A.

1

As defined by the USEPA.

2

Defined as twice the zero noise level by the USEPA.

M9110EH

Rev 0 5