Model 9110EH Instruction Manual Operating Instructions

Table 0-2: Test Functions Defined

Display Parameter Units

Description

RANGE RANGE

PPB, PPM,
UGM, MGM

The full scale limit to which the analyzer’s analog outputs are
currently set. This is not the physical range of the instrument. If
the Auto range mode has been selected, two RANGE functions
will appear. If the IND (independent) range mode has been
selected, three RANGE functions will appear:

NOX STB

STABILITY

PPB, PPM,
UGM, MGM

The stability is a standard deviation of the NO

X

concentration

over 25 samples, each recorded every 10 seconds. A low nox
stb
value indicates low variability in NO

X

.

SAMP FLW SAMPLE FLOW

cm³/min
(cc/m)

The flow rate of the sample gas through the reaction cell. This
value is not measured but calculated from the sample pressure.

OZONE FL

OZONE FLOW

cm³/min
(cc/m)

Flow rate of the O

3

gas stream as measured with a flow meter

PMT

PMT Signal

MV

The raw output voltage of the PMT.

NORM
PMT

NORMALIZED
PMT Signal

MV

The output voltage of the PMT after normalization for auto-zero
offset and temperature/pressure compensation (if activated).

AZERO AUTO-ZERO MV

The PMT signal with zero NO

X

,

which is usually slightly different

from 0 V. This offset is subtracted from the PMT signal and
adjusts for variations in the zero signal.

HVPS

HVPS

V

The PMT high voltage power supply.

RCELL
TEMP

REACTION CELL
TEMP

°C

The current temperature of the reaction cell.

BOX TEMP

BOX
TEMPERATURE

°C

The ambient temperature of the inside of the analyzer case.

PMT TEMP

PMT
TEMPERATURE

°C

The current temperature of the PMT.

IZS TEMP

IZS
TEMPERATURE

1

°C

The current temperature of the internal zero/span option. Only
appears when IZS option is enabled.

MOLY
TEMP

CONV
TEMPERATURE

°C

The current temperature of the NO

2

converter.

RCEL

REACTION CELL
PRESSURE

in-Hg-A

The current gas pressure of the reaction cell as measured at the
vacuum manifold. This is the vacuum pressure created by the
external pump.

SAMP

SAMPLE
PRESSURE

in-Hg-A

The current pressure of the sample gas as it enters the reaction
cell, measured between the NO/NO

x

and Auto-Zero valves.

NOX
SLOPE

NO

x

SLOPE

- -

The slope of the current NO

x

calibration as calculated from a

linear fit during the analyzer’s last zero/span calibration.

NOX OFFS NO

x

OFFSET

MV

The offset of the current NO

x

calibration as calculated from a

linear fit during the analyzer’s last zero/span calibration.

NO SLOPE NO SLOPE

- -

The slope of the current NO calibration as calculated from a
linear fit during the analyzer’s last zero/span calibration.

NO OFFS

NO OFFSET

MV

The offset of the current NO calibration as calculated from a
linear fit during the analyzer’s last zero/span calibration.

NO2

NO

2

concentration

PPB, PPM,
UGM, MGM

The current NO

2

concentration in the chosen unit.

NOX

NO

x

concentration

PPB, PPM,
UGM, MGM

The current NO

x

concentration in the chosen unit.

NO NO

concentration

PPB, PPM,
UGM, MGM

The current NO concentration in the chosen unit.

TEST TEST

SIGNAL

2

MV

Signal of a user-defined test function on output channel A4.

TIME

CLOCK TIME

hh:mm:ss

The current day time for iDAS records and calibration events.

M9110EH Rev 0