Figure 6-3, Setup for calibrating current outputs, Terminals (figure 0-3) – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual
Page 85
Model 9110EH Instruction Manual Operating Instructions
mA
IN OUT
V OUT +
V OUT -
I IN +
I IN -
Recording
Device
Analyzer
See Table 3-2 for
pin assignments of
the Analog Out
connector on the
rear panel.
Figure 0-3: Setup for Calibrating Current Outputs
CAUTION
Do not exceed 60 V between current loop outputs and instrument ground.
To adjust the zero and span values of the current outputs, press SETUP-MORE-DIAG-ENTR-
NEXT-NEXT and then:
EXAMPLE
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AIN A/C FREQUENCY: 60 HZ
SET> EDIT
EXIT
DIAG AIO
AOUT CALIBRATED: NO
< SET SET> CAL
EXIT
FROM ANALOG I/O CONFIGURATION MENU
DIAG AIO
CONC_OUT_2 RANGE: CURR
EXIT
DIAG AIO CONC_OUT_2 CALIBRATED: NO
< SET CAL
EXIT
DIAG AIO CONC_OUT_2 ZERO: 0 mV
U100 UP10 UP DOWN DN10 D100 ENTR EXIT
DIAG AIO
CONC_OUT_CURR, NO CAL
< SET SET> EDIT
EXIT
Press SET> to select the analog output channel
to be configured:. Then press EDIT to continue
DIAG AIO
AIN CALIBRATED: NO
SET> EDIT
EXIT
ENTR returns
to the previous
menu.
Increase or decrease
the current output by
100, 10 or 1 counts. The
resulting change in
output voltage is
displayed in the upper
line.
Continue adjustments
until the correct current
is measured with the
current meter.
DIAG AIO CONC_OUT_2 ZERO: 27 mV
U100 UP10 UP DOWN DN10 D100 ENTR EXIT
EXIT ignores the
new setting, ENTR
accepts the new
setting.
DIAG AIO CONC_OUT_2 CALIBRATED: YES
< SET CAL
EXIT
DIAG AIO CONC_OUT_2 SPAN: 10000 mV
U100 UP10 UP DOWN DN10 D100 ENTR EXIT
EXAMPLE
DIAG AIO CONC_OUT_2 ZERO: 9731 mV
U100 UP10 UP DOWN DN10 D100 ENTR EXIT
If a current meter is not available, an alternative method for calibrating the current loop
outputs is to connect a 250 Ω ±1% resistor across the current loop output. Using a
voltmeter, connected across the resistor, follow the procedure above but adjust the output
to the following values: