beautypg.com

Light leaks, Pneumatic operation, Pump and exhaust manifold – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 179

background image

Model 9110EH Instruction Manual Theory of Operation

10.1.5.3. Light Leaks

The M9110EH sensitivity curve includes a small portion of the visible light spectrum

(Figure 10-3), hence, it is important to make sure than the reaction cell is completely
sealed with respect to light. To ensure this, all pneumatic tubing leading into the reaction

cell is either opaque (vacuum exit tubing) in order to prevent light from entering the cell or
light penetration is prevented by stainless steel filters and orifices (gas entries).

10.2. Pneumatic Operation

CAUTION

It is important that the sample airflow system is leak-tight and not pressurized

over ambient pressure. Regular leak checks should be performed on the analyzer

as described in the maintenance schedule, Table 9-1. Procedures for correctly

performing leak checks can be found in Section 11.5.

10.2.1. Pump and Exhaust Manifold

NOTE

Relative Pressure versus absolute pressure. In this manual vacuum readings are

given in inches of mercury absolute pressure (in-Hg-A), i.e. indicate an absolute

pressure referenced against zero (a perfect vacuum).

The gas flow for the M9110EH is created by an external pump (Figure 10-5) that is pneumati-

cally connected through a 6.4 mm / 0.25” tube to the analyzer’s exhaust port located on
the rear panel (Figure 3-2). This pump creates a vacuum of approximately 5 in-Hg-A at one

standard liter/minute, which is provided to various pneumatic components by a vacuum
manifold located just in front of the rear panel (Figure 3-2 and Error! Reference source

not found.). Gas flow is created by keeping the analyzer’s sample gas inlet near ambient
pressure, usually by means of a small vent installed in the sample line at the inlet, in effect

pulling the gas through the instrument’s pneumatic systems.

There are several advantages to this external pump / pull-through configuration.

• By using an external pump, it is possible to remove a significant source of acoustic

noise and vibration from the immediate vicinity of the sensor. The PMT can act as a
“microphone”, amplifying noise and vibration within the chassis. This is one of the

main reasons, why the M9110EH has an external pump.

• Pumping heats and compresses the sample air, complicating the measurement

process if the pump is upstream.

• Most importantly, however, certain physical parts of the pump itself are made of

materials that might chemically react with the sample gas. Placing the pump
downstream of the reaction cell avoids these problems.

See Section 3.1 for information on connecting and operating the external pump.

M9110EH Rev 0

165