Analog output offset adjustment – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual
Page 83
Model 9110EH Instruction Manual Operating Instructions
Press SET> to select the analog output channel to be configured:
DISPLAYED
AS
=
CHANNEL
CONC_OUT_1
=
A1
CONC_OUT_2
=
A2
CONC_OUT_3
=
A3
TEST
OUTPUT
=
A4
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AOUTS CALIBRATED: NO
< SET SET> CAL
EXIT
FROM ANALOG I/O CONFIGURATION MENU
DIAG AIO
CONC_OUT_1 :5V, NO CAL
< SET SET> EDIT
EXIT
DIAG AIO CONC_OUT_1 CALIBRATED: YES
< SET CAL
EXIT
DIAG AIO
CONC_OUT_1 RANGE: 5V
SET> EDIT
EXIT
DIAG AIO CONC_OUT_2 CALIBRATED: NO
< SET CAL
EXIT
DIAG AIO CONC_OUT_1 VOLT–Z : 0 mV
U100 UP10 UP DOWN DN10 D100 ENTR EXIT
These keys increase / decrease the analog
output by 100, 10 or 1 counts. Continue
adjustments until the voltage measured at the
output of the analyzer and/or the input of the
recording device matches the value in the upper
right hand corner of the display to the tolerance
listed in Table 6-10.
The concentration display will not change. Only
the voltage reading of your voltmeter will change.
DIAG AIO CONC_OUT_1 VOLT–S : 4500 mV
U100 UP10 UP DOWN DN10 D100 ENTR EXIT
DIAG AIO
CONC_OUT_1 REC OFS: 0 mV
< SET SET> EDIT
EXIT
If AutoCal is ON, go to
Section 6.7.3
EXIT ignores the
new setting.
ENTR accepts the
new setting.
DIAG AIO CONC_OUT_1 AUTO CAL: OFF
< SET SET> EDIT
EXIT
6.7.3.4. Analog Output Offset Adjustment
Some analog signal recorders require that the zero signal is significantly different from the
baseline of the recorder in order to record slightly negative readings from noise around the
zero point. This can be achieved in the M200E by defining a zero offset, a small voltage
(e.g., 10% of span), which can be added to the signal of individual output channels just like
a strip chart recorder can be set to have a 10% offset by moving the pens. Set the offset by
pressing SETUP-MORE-DIAG-ENTR-NEXT-NEXT and the following keys:
M9110EH Rev 0
69