beautypg.com

Auto range mode (auto), Range units, And no – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 70: Mols per mol, ppb), parts per million (10, Ug) or milligrams per cubic meter (mg/m

background image

Operating Instructions Model 9110EH Instruction Manual

6.5.5. Auto Range Mode (AUTO)

In AUTO range mode, the analyzer automatically switches the reporting range between two
user-defined ranges (low and high). The same low and high span settings are applied

equally to NO, NO

2

and NO

X

readings. The unit will switch from low range to high range

when either the NO, or NO

X

concentration exceeds 98% of the low range span. The unit will

return from high range back to low range once both the NO and NO

X

concentrations fall

below 75% of the low range span.

In AUTO range mode the RANGE test function displayed on the front panel will be replaced

by two separate functions:

RANGE1: The LOW range setting for all analog outputs.
RANGE2: The HIGH range setting for all analog outputs.

The high/low range status is also reported through the external, digital status bits

(Section 6.9.1). To set individual ranges press the following keystroke sequence.



SAMPLE RANGE = 500.0 PPB NOX=X.X

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS RNGE PASS CLK MORE

EXIT

SETUP X.X

RANGE CONTROL MENU

MODE

SET UNIT

EXIT

SETUP X.X

RANGE MODE: SNGL

SNGL DUAL AUTO

ENTR EXIT

EXIT x 2 returns

to the main

SAMPLE display

SETUP X.X

RANGE MODE: AUTO

SNGL

DUAL AUTO

ENTR EXIT

SETUP X.X

RANGE CONTROL MENU

MODE

SET UNIT

EXIT

SETUP X.X

HIGH RANGE: 500.0 Conc

0 0 5 0 0 .0 ENTR EXIT

SETUP X.X

LOW RANGE: 500.0 Conc

0 0 5 0 0 .0 ENTR EXIT

Toggle the numeral

keys to set the

LOW and HIGH

range value.

ENTR accepts the

new setting, EXIT

ignores the new

setting.

NOTE

Low and high ranges have separate slopes and offsets for computing the NO and

NOx concentrations. Hence, the two ranges must be calibrated independently.

6.5.6. Range Units

The M9110EH can display concentrations in parts per billion (10

9

mols per mol, PPB), parts per

million (10

6

mols per mol, PPM), micrograms per cubic meter (µg/m

3

, UG) or milligrams per

cubic meter (mg/m

3

, MG). Changing units affects all of the display, COM port and iDAS

values for all reporting ranges regardless of the analyzer’s range mode. To change the
concentration units:

56 M9110EH Rev 0