beautypg.com

Warning messages – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 36

background image

Getting Started Model 9110EH Instruction Manual

Name Color Behavior

Significance

STATUS LEDs

Sample Green On

Unit is operating in sample mode, front panel
display is continuously updated.

Cal

Yellow

Off

The instrument’s calibration is not enabled.

Fault Red Blinking The analyzer is warming up and out of

specification for a fault-free reading.

3.5.3. Warning Messages

During warm-up, internal temperatures and other parameters may be outside of specified
limits. The software will suppress most warning conditions for 30 minutes after power up.

The following table includes a brief description of the various warning messages that may
appear after the warm-up time. If warning messages persist after 30 minutes, investigate
their cause using the troubleshooting guidelines in Chapter 11. To view and clear warning

messages, use the following key strokes:

SAMPLE

HVPS WARNING NOX = 0.0


TEST CAL

MSG CLR SETUP

Press CLR to clear the current

message.

If more than one warning is active, the

next message will take its place
Once the last warning has been

cleared, the analyzer returns to

SAMPLE mode

SAMPLE

RANGE=500.0 PPB NO = 0.0


< TST TST > CAL

MSG CLR SETUP

SAMPLE

HVPS WARNING NOX = 0.0


TEST CAL

MSG CLR SETUP

TEST deactivates warning

messages

MSG activates warning

messages.

> keys replaced with

TEST key

NOTE:

If the warning message persists after several attempts to

clear it, the message may indicate a real problem and not

an artifact of the warm-up period

22 M9110EH Rev 0