beautypg.com

Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 160

background image

Instrument Maintenance Model 9110EH Instruction Manual

Item Action

Frequency

Cal

Check ?

Manual

Section

Date Performed

NO

2

converter

Check
conversion

efficiency

Every 3
months

Yes on
change,

else no

9.3.8

External zero
air scrubber

(optional)

Exchange
chemical

Every 3
months

No 9.3.7

External dryer

(optional)

Replace

chemical

When

indicator
color

changes

No

Ozone filter

Change

chemical

Annually

Yes 9.3.3

Reaction cell
window

Clean Annually

or

as

necessary

Yes 9.3.9

DFU filter (air

inlet filter of
Perma Pure

dryer, pump

dust filter
and IZS

filter)

Change

particle filter

Annually No

9.3.2
9.3.7

Pneumatic

sub-system

Check for leaks

in gas flow
paths

Annually or

after
repairs

involving

pneumatics

Yes on

leaks,
else no

11.5.1,

11.5.2

Reaction cell
and O

3

O-

rings and

sintered
filters

Replace Annually

Yes

on

Rcell or

PMT O-

rings.

9.3.10

PMT Sensor
Hardware

Calibration

Low-level
hardware

calibration

On PMT/
preamp

changes or
if slope is

outside of

1.0±0.3

Yes

7.9

Pump

Rebuild head

when RCEL
pressure

exceeds 10

in-Hg-A (at

sea level)

Yes 9.3.4

NO

2

converter

Replace
converter

Every 3
years or if

conversion

efficiency
drops below

96%

Yes (if

CE factor

is used)

9.3.8

Items shown in bold face are required to maintain full warranty, all other items are strongly recommended.

146 M9110EH Rev 0