Specifications, Approvals and Warranty Model 9110EH Instruction Manual

2.2. EPA Equivalency Designation

The Model 9110EH Analyzer is designated as Reference Method Number RFNA-1194-099

(same designation as model M9110EH) as per 40 CFR Part 53 when operated under the
following conditions:

• Range: Any full-scale range between 0-0.05 and 0-1.0 ppm (parts per million).
• Ambient temperature range of 5

to 40

o

C.

• Line voltage range of 105-125 VAC or 220-240 VAC, at 50 or 60 Hz.
• Equipped with PTFE filter element in the internal filter assembly.
• Equipped with ozone supply air filter

External vacuum pump capable of 10 in-Hg-A at 2 standard liters per minute (slpm) or
better.

Software settings:

Dynamic span

OFF

Dynamic zero

OFF

Cal-on-NO

2

OFF

Dilution factor

OFF

Temp/Pressure compensation

ON

AutoCal

ON or OFF

Independent range

ON or OFF

Auto-range

ON or OFF

Converter efficiency

Acceptable values of 0.96 to 1.02


Under the designation, the analyzer may be operated with or without the following optional
equipment:

• Rack mount with or without slides.
• Rack mount for external pump.
• Zero/span valve options.
• Nafion-type permeation sample gas conditioner
• Internal zero/span (IZS) option with:

• NO

2

permeation tube - 0.4ppm at 0.7 liter per minute; certified/uncertified.

• NO

2

permeation tube - 0.8 ppm at 0.7 liter per minute; certified/uncertified.

Under the designation, the IZS option cannot be used as the source of calibration.

• 4-20mA isolated analog outputs.
• Status outputs.
• Control inputs.
• RS-232 output.
• Ethernet communications option

6 M9110EH Rev 0