2.2. EPA Equivalency Designation
The Model 9110EH Analyzer is designated as Reference Method Number RFNA-1194-099
(same designation as model M9110EH) as per 40 CFR Part 53 when operated under the
following conditions:
• Range: Any full-scale range between 0-0.05 and 0-1.0 ppm (parts per million).
• Ambient temperature range of 5
to 40
o
C.
• Line voltage range of 105-125 VAC or 220-240 VAC, at 50 or 60 Hz.
• Equipped with PTFE filter element in the internal filter assembly.
• Equipped with ozone supply air filter
External vacuum pump capable of 10 in-Hg-A at 2 standard liters per minute (slpm) or
better.
Software settings:
Dynamic span
OFF
Dynamic zero
OFF
Cal-on-NO
2
OFF
Dilution factor
OFF
Temp/Pressure compensation
ON
AutoCal
ON or OFF
Independent range
ON or OFF
Auto-range
ON or OFF
Converter efficiency
Acceptable values of 0.96 to 1.02
Under the designation, the analyzer may be operated with or without the following optional
equipment:
• Rack mount with or without slides.
• Rack mount for external pump.
• Zero/span valve options.
• Nafion-type permeation sample gas conditioner
• Internal zero/span (IZS) option with:
• NO
2
permeation tube - 0.4ppm at 0.7 liter per minute; certified/uncertified.
• NO
2
permeation tube - 0.8 ppm at 0.7 liter per minute; certified/uncertified.
Under the designation, the IZS option cannot be used as the source of calibration.
• 4-20mA isolated analog outputs.
• Status outputs.
• Control inputs.
• RS-232 output.
• Ethernet communications option