Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual
Page 265
Model 9110EH Instruction Manual APPENDIX A
Name Description
Units
NOZSC2
NO concentration for range #2 during zero/span calibration,
just before computing new slope and offset
ppb
2
N2ZSC1 NO
2
concentration for range #1 during zero/span calibration,
just before computing new slope and offset
ppb
2
N2ZSC2 NO
2
concentration for range #2 during zero/span calibration,
just before computing new slope and offset
ppb
2
NXCNC1 NO
X
concentration for range #1
ppb
2
NXCNC2 NO
X
concentration for range #2
ppb
2
NOCNC1
NO concentration for range #1
ppb
2
NOCNC2
NO concentration for range #2
ppb
2
N2CNC1 NO
2
concentration for range #1
ppb
2
N2CNC2 NO
2
concentration for range #2
ppb
2
STABIL Concentration
stability
ppb
2
AZERO
Auto zero offset (range de-normalized)
mV
O3FLOW
Ozone flow rate
cm³/min
RCPRES Reaction
cell
pressure
in-Hg-A
RCTEMP Reaction
cell
temperature
°C
MFTEMP
1
Bypass or dilution manifold temperature
°C
IZTEMP
1
IZS block temperature
°C
CNVEF1
Converter efficiency factor for range #1
—
CNVEF2
Converter efficiency factor for range #2
—
CNVTMP Converter
temperature
°C
PMTTMP PMT
temperature
°C
SMPFLW Sample
flow
rate
cm³/min
SMPPRS Sample
pressure
in-Hg-A
BOXTMP Internal
box
temperature
°C
HVPS
High voltage power supply output
V
REFGND
Ground reference (REF_GND)
mV
RF4096
4096 mV reference (REF_4096_MV)
mV
TEST11
Diagnostic test input (TEST_INPUT_11)
mV
TEST13
Diagnostic test input (TEST_INPUT_13)
mV
TEMP5
Diagnostic temperature input (TEMP_INPUT_5)
°C
TEMP6
Diagnostic temperature input (TEMP_INPUT_6)
°C
1
Factory
option.
2
Standard instrument units.
M9110EH Rev 0
251