Model 9110EH Instruction Manual Operating Instructions

Table 0-22: M9110EH Default iDAS Configuration, Revision C.7.

List of Channels

List of Parameters

NAME – MODE/EVENT - PRECISION - NUM SAMPLES

NOXCNC1 - AVG - 4 - ON

NOCNC1 - AVG - 4 - OFF

N2CNC1 - AVG - 4 - OFF

Name: CONC

Event: ATIMER

Sample Period: 000:00:01

Report Period: 000:01:00

Number of Records: 800

RS-232 report: OFF
Channel enabled: ON

DAS HOLDOFF: ON

STABIL - AVG - 4 - OM

NXZSC1 - SLPCHG - 4 - OFF
NOXSLP1 - SLPCHG - 4 - OFF
NOXOFFS1 - SLPCHG - 4 - OFF
NOZSC1 - SLPCHG - 4 - OFF
NOSLP1 - SLPCHG - 4 - OFF
NOOFFS1 - SLPCHG - 4 - OFF
N2ZSC1 - SLPCHG - 4 - OFF
CNVEF1 - SLPCHG - 4 - OFF

Name: CALDAT

Event: SLPCHG
Number of Records: 200

RS-232 report: OFF

Channel enabled: ON

DAS HOLDOFF: OFF

STABIL - SLPCHG - 4 - OFF
NXZSC1 - EXITMP - 4 - OFF

NOZSC1 - EXITMP - 4 - OFF

N2ZSC1 - EXITMP - 4 - OFF

Name: CALCHECK

Event: EXITMP

Number of Records: 200

RS-232 report: OFF

Channel enabled: ON
DAS HOLDOFF: OFF

STABIL - EXITMP - 4 - OFF
SMPFLW - AVG – 2 - OFF
O3FLOW – AVG – 2 - OFF
RCPRESS – AVG – 2 - OFF
SMPPRES – AVG – 2 - OFF
RCTEMP - AVG – 2 - OFF
PMTTMP - AVG – 2 - OFF
CNVTMP - AVG – 2 - OFF
BOXTMP - AVG – 2 - OFF
HVPS - AVG – 2 - OFF

Name: DIAG

Event: ATIMER

Sample Period: 000:00:30

Report Period: 001:00:00
Number of Records: 1100

RS-232 report: OFF

Channel enabled: ON

DAS HOLDOFF: OFF

AZERO - AVG – 2 - OFF
NOXCNC1 - AVG - 4 - OFF
NOCNC1 – AVG - 4 - OFF
N2CNC1 - AVG – 4 - OFF
STABIL - AVG – 4 - OFF
SMPFLW - AVG – 2 - OFF
O3FLOW - AVG – 2 - OFF
RCPRESS - AVG – 2 - OFF
SMPPRES - AVG – 2 - OFF
RCTEMP - AVG – 2 - OFF
PMTTMP - AVG – 2 - OFF
CNVTMP - AVG – 2 - OFF
BOXTMP - AVG – 2 - OFF
HVPS - AVG – 1 - OFF
AZERO - AVG – 2 - OFF
REFGND - AVG – 1 - OFF

Name: HIRES

Event: ATIMER

Sample Period: 000:00:01

Report Period: 000:00:01
Number of Records: 1500

RS-232 report: OFF

Channel enabled: OFF

DAS HOLDOFF: OFF

REF4096 - AVG - 1 - OFF

DIAG: Daily averages of temperature zones, flow and pressure data as well as some other
diagnostic parameters (HVPS, AZERO). These data are intended for predictive diagnostics

and maintenance described in Chapter 9. The last 1100 daily averages are stored to cover
more than four years of analyzer performance, a major portion of the analyzer’s lifetime. If

these parameters are required for data analysis, this channel can easily be duplicated (e.g.,

M9110EH Rev 0