APPENDIX A Model 9110EH Instruction Manual

Setup Variable

Numeric

Units

Default

Value

Value

Range

Description

OFLOW_SET

cc/m

80

0–500

Ozone flow warning limits. Set
point is not used.

OZONE_FLOW_SLOPE —

1

0.001-
100

Slope term to correct ozone flow
rate.

RCELL_SAMP_RATIO

0.53

0.1-2

Maximum reaction cell pressure /

sample pressure ratio for valid
sample flow calculation.

STD_BOX_TEMP ºK

298

Valid limits:
278-338

1-500

Standard box temperature and

valid limits for temperature
compensation.

STD_RCELL_TEMP ºK

323

Valid limits:
278-338

1-500

Standard reaction cell temperature

and valid limits for temperature
compensation.

STD_RCELL_PRESS "Hg

7

Valid limits:
0.5-12

0.1-50

Standard reaction cell pressure

and valid limits for pressure
compensation.

STD_SAMP_PRESS "Hg

29.92
Valid limits:

0.5-32

0.1-50

Standard sample pressure and
valid limits for pressure

compensation.

PRESS_FILT_SIZE

5

Samples

60

1-120

Sample and reaction cell pressure
moving average filter size.

PRESS_SAMP_FREQ

5

Minutes

2

0.1-60

Sample and reaction cell pressure

sampling frequency.

RS232_MODE

0

0-65535

RS-232 COM1 mode flags. Add

values to combine flags.
1 = quiet mode

2 = computer mode
4 = enable security

8 = enable iChip support

16 = enable Hessen protocol
32 = enable multidrop

64 = enable modem

128 = ignore RS-232 line errors
256 = disable XON / XOFF support

1024 = enable RS-485 mode

2048=even parity, 7 data bits, 1

stop bit
4096 = enable command prompt

BAUD_RATE — 19200

300,

1200,
2400,

4800,

9600,
19200,

38400,

57600,
115200

RS-232 COM1 baud rate. Enclose

value in double quotes (") when
setting from the RS-232 interface.

240 M9110EH Rev 0