9.2. Predictive Diagnostics

The analyzer’s test functions can be used to predict failures by looking at trends in their

values. Initially it may be useful to compare the state of these test functions to the values
measured on your instrument at the factory and recorded on the M9110EH Final Test and

Validation Data Form attached to the manual). Table 9-2 can be
used as a basis for taking action as these values change with time. The internal data
acquisition system (iDAS) is a convenient way to record and track these changes. APICOM

control software can be used (Section 6.11.5 describes APICOM).

Table 9-2: Predictive Uses for Test Functions

Function

Expected

Actual

Interpretation & Action

Fluctuating

Developing leak in pneumatic system. Check for
leaks

RCEL
pressure

Constant to
within ± 0.5

Slowly increasing

Pump performance is degrading. Replace pump
head when pressure is above 10 in-Hg-A

Fluctuating

Developing leak in pneumatic system. Check for

leaks

Slowly increasing

Flow path is clogging up. Replace orifice filters

SAMPLE

pressure

Constant

within
atmospheric

changes

Slowly decreasing

Developing leak in pneumatic system to vacuum
(developing valve failure). Check for leaks

Ozone
Flow

Constant to
within ± 15

Slowly decreasing

Flow path is clogging up. Replace orifice filters

Developing AZERO valve failure. Replace valve
PMT cooler failure. Check cooler, circuit, and
power supplies
Developing light leak. Leak check.

AZERO

Constant

within ±20 of
check-out

value

Significantly

increasing

O

3

air filter cartridge is exhausted. Change

chemical

NO2
CONC

Constant for
constant

concentrations

Slowly decreasing
signal for same

concentration

Converter efficiency may be degrading. Replace
converter.

Change in instrument response. Low level
(hardware) calibrate the sensor

NO2
CONC

(IZS)

Constant
response from

day to day

Decreasing over
time

Degradation of IZS permeation tube. Change
permeation tube

NO2

CONC

(IZS)

Constant

response from

day to day

Heavily fluctuating

from day to day

Ambient changes in moisture are affecting the

performance. Add a dryer to the zero air inlet.

NO CONC

Constant for
constant

concentration

Decreasing over
time

Drift of instrument response; clean RCEL
window, change O

3

air filter chemical.

9.3. Maintenance Procedures

The following procedures need to be performed regularly as part of the standard

maintenance of the Model 9110EH.