beautypg.com

Turn on the cal_on_no, Feature – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 136

background image

Calibration Procedures Model 9110EH Instruction Manual

Wait until STABIL

value falls below

0.5 ppb. This may

take several

minutes.

ZERO CAL M NOX STB=XXX.X PPB NOX=X.X

< TST TST > ZERO CONC

ZERO CAL M NOX STB=XXX.X PPB NOX=X.X

< TST TST > ENTR CONC

EXIT

If the ZERO button does
not appear, see Chapter

11 for troubleshooting.

Analyzer enters ZERO CAL

mode

EXIT terminates

calibration and returns to

SAMPLE mode.

ZERO CAL M NOX STB=XXX.X PPB NOX=X.X

< TST TST > ZERO CONC

EXIT

ZERO CAL M NOX STB=XXX.X PPB NOX=X.X X

< TST TST > CAL CALZ CALS

SETUP

SAMPLE RANGE = 500.0 PPB NOX= X.X

< TST TST > CAL CALZ CALS

SETUP

SAMPLE NOX STB=XXX.X PPB NO = X.X

< TST TST > CAL CALZ CALS

SETUP

Set the Display to show the

NOX STB test function.

Pressing ENTR changes the
calibration of the instrument.

Turn ON the CAL_ON_NO

2

feature:

SAMPLE RANGE = 500.0 PPB NOX=X.X

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS RNGE PASS CLK MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM VARS DIAG

EXIT

SETUP X.X

0 ) DAS_HOLD_OFF=15.0 Minutes

NEXT JUMP

EDIT PRNT EXIT

SETUP X.X

ENTER VARS PASS: 818

8 1 8

ENTR EXIT

EXIT ignores the new setting.

ENTR accepts the new setting.

SETUP X.X

8) CAL_ON_NO

2

= OFF

PREV NEXT JUMP

EDIT PRNT EXIT

SETUP X.X

8) CAL_ON_NO

2

= ON


ON

ENTR EXIT

Toggle this key to change setting

Press NEXT until …

122 M9110EH Rev 0