beautypg.com

Dilution ratio – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 71

background image

Model 9100EH Instruction Manual Operating Instructions

SAMPLE RANGE = 500.0 PPB NOX=X.X

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS RNGE PASS CLK MORE

EXIT

SETUP X.X

RANGE CONTROL MENU

MODE

SET UNIT

EXIT

SETUP X.X CONC UNITS: PPM

PPM PPB UGM MGM

ENTER EXIT

Select the preferred

concentration unit.

SETUP X.X CONC UNITS: UGM

PPM PPB UGM MGM

ENTER EXIT

EXIT returns

to the main

menu.

ENTR accepts

the new unit,

EXIT returns

to the SETUP

menu.

Conversion factors from volumetric to mass units used in the M9100EH:

NO: ppb x 1.34 = µg/m

3

; ppm x 1.34 = mg/m

3

NO

2

: ppb x 2.05 = µg/m

3

; ppm x 2.05 = mg/m

3

Concentrations displayed in mg/m

3

and µg/m

3

use 0° C and 760 Torr as standard

temperature and pressure (STP). Consult your local regulations for the STP used by your

agency. EPA protocol applications, for example, use 25° C as the reference temperature.
Changing the units may cause a bias in the measurements if standard temperature and
pressure other than 0°C and 760 Torr are used. This problem can be avoided by recalibrat-

ing the analyzer after any change from a volumetric to a mass unit or vice versa.

CAUTION

In order to avoid a reference temperature bias, the analyzer must be recalibrated

after every change in reporting units.

6.5.7. Dilution Ratio

The dilution ratio is a software option that allows the user to compensate for any dilution of

the sample gas before it enters the sample inlet. Using the dilution ratio option is a 4-step
process:

• Select concentration units: Follow the procedure in Section 6.5.6.

M9110EH Rev 0

57