APPENDIX A Model 9110EH Instruction Manual
Setup Variable
Numeric
Units
Default
Value
Value
Range
Description
TEST_CHAN_ID —
NONE NONE,
PMT DE-
TECTOR,
OZONE
FLOW,
SAMPLE
FLOW,
SAMPLE
PRESSUR
E,
RCELL
PRESSUR
E,
RCELL
TEMP,
MANIFOL
D TEMP,
IZS
TEMP,
CONV
TEMP,
PMT
TEMP,
CHASSIS
TEMP,
HVPS
VOLTAGE
Diagnostic analog output ID.
Enclose value in double quotes (")
when setting from the RS-232
interface.
COMMAND_PROMPT —
“Cmd>
”
Any
character
in the
allowed
character
set. Up
to 100
character
s long.
RS-232 interface command
prompt. Displayed only if enabled
with RS232_MODE variable.
Enclose value in double quotes (")
when setting from the RS-232
interface.
REMOTE_CAL_MODE
3
—
LOW
LOW,
HIGH
Range to calibrate during remote
calibration. Enclose value in double
quotes (") when setting from the
RS-232 interface.
PASS_ENABLE
—
OFF
ON, OFF
ON enables passwords; OFF
disables them.
STABIL_GAS — NOX NO,
NO2,
NOX
Selects gas for stability
measurement. Enclose value in
double quotes (") when setting
from the RS-232 interface.
STABIL_FREQ
Seconds
10
1-300
Stability measurement sampling
frequency.
STABIL_SAMPLES
Samples
25
2-40
Number of samples in concentra-
tion stability reading.
HVPS_SET Volts
650
Warnings:
400-900
0-2000
High voltage power supply warning
limits. Set point is not used.
242 M9110EH Rev 0