Getting Started Model 9110EH Instruction Manual

Set the expected NO and NOx span gas concentration. If you supply NO gas to the

analyzer, the values for expected NO and NO

x

need to be identical.

This sequence causes the

analyzer to prompt for the

expected NO and NO

x

span concentration.

EXIT ignores the new setting and

returns to the CONCENTRATION

MENU.

ENTR accepts the new setting and

returns to the

CONCENTRATION MENU..

M-P CAL CONCENTRATION MENU

NOX NO CONV

EXIT

M-P CAL NOX SPAN CONC: 400.0 Conc

0 0 0 4 0 .0 ENTR EXIT

M-P CAL CONCENTRATION MENU

NOX NO CONV

EXIT

M-P CAL NO SPAN CONC: 400.0 Conc

0 0 0 4 0 .0 ENTR EXIT

M-P CAL RANGE = 500.0 PPB NO = X.X

< TST TST > ZERO CONC

EXIT

SAMPLE RANGE = 500.0 PPB NOX= X.X

< TST TST > CAL SETUP

The NO

x

& NO span concentration

values automatically default to

400.0 Conc.

To change this value to the actual

concentration of the span gas, enter

the number by pressing the key

under each digit until the expected

value appears.

28 M9110EH Rev 0