Warm-up, Figure 3-10, Front panel display during startup sequence – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 35: Table 3-5, Front panel display during system warm-up, Display as shown in figure 3-10. the an

background image

Model 9110EH Instruction Manual Getting Started

Firmware

fully booted

The revision level of the firmware

installed in your analyzer is briefly

displayed

SOFTWARE REVISION C.4

BOOT PROGRESS [XXXXXXXX 80% _ _]

The instrument is loading the

analyzer firmware.

M200E NOX ANALYZER

BOOT PROGRESS [XXXXX 50%_ _ _ _ _]

The instrument is loading

configuration and calibration data

from the flash chip

STARTING INSTRUMENT W/FLASH : 1

System waits 3 seconds then

automatically begins its initialization

routine. No action required.

System is checking the format of

the instrument’s flash memory chip.

SELECT DESIRED FUNCTION : 3

START

.

CHECKING FLASH STATUS

: 2

STARTING INSTRUMENT CODE : 1

SAMPLE SYSTEM RESET NOX=X.X

TEST CAL CLR SETUP

If at this point,

**FLASH FORMAT INVALID**
appears, contact T–API customer

service

Press CLR to clear initial warning messages.

(see Section 3.2.3)

Figure 3-10: Front Panel Display During Startup Sequence

3.5.2. Warm-Up

The M9110EH requires about 30 minutes warm-up time before reliable NO

X

, NO and NO

2

measurements can be taken. During that time, various portions of the instrument’s front

panel will behave as follows. See Figure 3-4 for locations.

Table 3-5: Front Panel Display During System Warm-Up

Name Color Behavior

Significance

Concentra-
tion Field

N/A

Switches between
NO

X

, NO and NO

2

This is normal operation.

Mode Field

N/A

Displays blinking
“SAMPLE”

Instrument is in sample mode but is still in the
process of warming up (hold-off period is

active).

M9110EH

Rev 0 21