Troubleshooting & Repair Model 9110EH Instruction Manual

Figure 11-4: Relay Board PCA

To enter the signal I/O test mode to manually control I/O functions such as valves and

heaters, press the following keys while observing the relay board LEDs when toggling:

=

SAMPLE RANGE = 500.0 PPB NOX=X.X

< TST TST > CAL

SETUP

EXAMPLE

SETUP X.X

PRIMARY SETUP MENU


CFG DAS RNGE PASS CLK MORE

EXIT

SETUP X.X SECONDARY SETUP MENU

COMM VARS DIAG

EXIT

DIAG

SIGNAL I / O


PREV NEXT
JUMP

ENTR

EXIT

DIAG I / O Test Signals Displayed Here

PREV NEXT JUMP

PRNT EXIT

Use the JUMP key to

go directly to a

specific signal

See Appendix A-4 for

a complete list of

available SIGNALS

DIAG I / O

JUMP TO: 5


0 5

ENTR EXIT

Enter 05 to Jump

to Signal 5:

(CAL_LED)

DIAG I / O CAL_LED = ON

PREV NEXT JUMP

ON PRNT EXIT

Exit to return

to the

DIAG menu

SETUP X.X

ENTER DIAG PASS: 818


8 1 8

ENTR EXIT

Use the NEXT & PREV

keys to move between

signal types.

Pressing the PRNT key will send a formatted printout to the serial port and can be

captured with a computer or other output device.

EXIT

returns

to the

main

SAMPLE

menu