Model 9110EH Instruction Manual Operating Instructions

Setting analog range limits to different values does not affect the instrument’s calibration.

To select the IND range mode, press the following keys:

SAMPLE RANGE = 500.0 PPB NOX=X.X

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU


CFG DAS RNGE PASS CLK MORE

EXIT

SETUP X.X

RANGE CONTROL MENU

MODE

SET UNIT

EXIT

SETUP X.X

RANGE MODE: SNGL

SNGL IND AUTO

ENTR EXIT

SETUP X.X

RANGE MODE: IND

SNGL

DUAL AUTO

ENTR EXIT

EXIT ignores the new setting and returns to the

RANGE CONTROL MENU.

ENTR accepts the new setting and returns to the

RANGE CONTROL MENU.

To set the range value for each independent range, press:



SETUP X.X

RANGE CONTROL MENU

MODE

SET UNIT

EXIT

SETUP X.X NO RANGE: 500.0 Conc

0 1 0 0 0 .0 ENTR EXIT

SETUP X.X NOX RANGE: 500.0 Conc

0 1 5 0 0 .0 ENTR EXIT

SETUP X.X NO2 RANGE: 500.0 Conc

0 0 5 0 0 .0 ENTR EXIT

Press the

number keys to
set range value

EXIT ignores the new

setting and returns to the

RANGE CONTROL

MENU.

ENTR accepts the new

setting and returns to the

RANGE CONTROL

MENU.

SAMPLE RANGE = 500.0 PPB NOX=X.X

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS RNGE PASS CLK MORE

EXIT

M9110EH Rev 0