beautypg.com

Calibration checks with izs or zero/span valves, Figure 7-4, 9110eh – Teledyne 9110EH - Nitrogen Oxides Analyzer User Manual

Page 137

background image

Model 9110EH Instruction Manual Calibration Procedures

Perform the span calibration procedure:

If the SPAN button does

not appear, see Section

11 for troubleshooting.

SPAN CAL M NOX STB=XXX.X PPB NOX=X.X

< TST TST > ENTR CONC

EXIT

EXIT terminates the

calibration process and

returns the unit to

SAMPLE mode.

SPAN CAL M NOX STB=XXX.X PPB NOX=X.X

< TST TST > SPAN CONC

EXIT

SAMPLE RANGE = 500.0 PPB NOX= X.X

< TST TST > CAL CALZ CALS

SETUP

Set the Display to show the NOX

STB test function.

Analyzer enters SPAN CAL

mode.

SPAN CAL M NOX STB=XXX.X PPB NOX=X.X

< TST TST > SPAN CONC

EXIT

Pressing ENTR changes the
calibration of the instrument.

EXIT returns to the

SAMPLE mode.

Wait until the STABIL value

falls below 0.5 ppb. This

may take several minutes.

SAMPLE NOX STB=XXX.X PPB NOX=X.X X

< TST TST > CAL CALZ CALS

SETUP

Turn off the CAL_ON_NO2 feature (step four above).

7.6. Calibration Checks with IZS or Zero/Span

Valves

To perform a manual calibration check with zero/span valve or IZS option installed:

• Connect the sources of Zero Air and Span Gas as shown below.

Figure 7-4: Setup for Manual Calibration Check with Z/S Valve or IZS Option

M9110EH Rev 0

123

9110EH