Operating Instructions Model 9110EH Instruction Manual

6.7.5. Optic Test

The optic test function tests the response of the PMT sensor by turning on an LED located in
the cooling block of the PMT (Fig. 10-15). The analyzer uses the light emitted from the LED

to test its photo-electronic subsystem, including the PMT and the current to voltage
converter on the pre-amplifier board. To make sure that the analyzer measures only the

light coming from the LED, the analyzer should be supplied with zero air. The optic test
should produce a PMT signal of about 2000±1000 mV. To activate the electrical test press
the following key sequence.

SAMPLE RANGE = 500.0 PPB NOX=X.X

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU


CFG DAS RNGE PASS CLK MORE

EXIT

SETUP X.X SECONDARY SETUP MENU

COMM VARS DIAG

EXIT

DIAG

SIGNAL I / O


PREV NEXT
JUMP

ENTR

EXIT

DIAG OPTIC RANGE = 500.0 PPB NOX=X.X

EXIT

SETUP X.X ENTER DIAG PASS: 818

8 1 8

ENTR EXIT

Press NEXT until…

DIAG

OPTIC TEST


PREV NEXT

ENTR EXIT

Press TST until…

DIAG ELEC PMT = 2751 MV NOX=X.X


EXIT

While the optic test is

activated, PMT should be

2000 mV ± 1000 mV

NOTE

This is a coarse test for functionality and not an accurate calibration tool. The

resulting PMT signal can vary significantly over time and also changes with low-

level calibration.

6.7.6. Electrical Test

The electrical test function creates a current, which substitutes the PMT signal, and feeds it
into the preamplifier board. This signal is generated by circuitry on the pre-amplifier board

itself and tests the filtering and amplification functions of that assembly along with the A/D
converter on the motherboard. It does not test the PMT itself. The electrical test should
produce a PMT signal of about 2000 ±1000 mV.

74 M9110EH Rev 0