Appendix F. Calibration and Zeroing

The currently measured value for the sensor will be displayed in the next

screen. Now place the sensor into the first known condition, and enter that

known value into the

First calibrated value box. Press Set First Value. Wait

for the calibration process to measure the first value. The word Calibrating

will be visible in the

Current Value box until that process is complete. Now

place the sensor into the second known condition, and then enter the

corresponding known value into the

Second calibrated value box. Press Set

Second Value. The calibration process measures the second point value. At

that point the datalogger calculates the new multiplier and offset and applies

them within the running program. These values are also written to the

calibration file.

F-10