Operator's Guide

adc­command module calib­noise arguments

Perform a broad-band noise calibration according to the
specified arguments, which are given as a space-separated

list of key=value pairs:

component=[ ALL | Z | N/S | E/W ] - specify

the component(s) to be calibrated

duration=m - specify the duration, in minutes, where
m is an integer value. For accurate results, the

duration should be significantly longer than the
response time of the instrument.

amplitude=a - specify the amplitude, where a is the

integer percentage of the full output of the cailbration
signal generator. This can normally be given as 100

but should be reduced if clipping is noticed.

adc­command module obs­deploy

Deploy an OBS system by levelling the bowl, then unlocking

and centring each mass in turn. This sub-command takes no
options.

adc­command module obs­level bowl=which

Level the bowl or bowls of an OBS system. The parameter
which should be one of 1BOWL (for single-bowl units), Z, N/S 

or E/W.

adc­command module obs­datum bowl=which

Return the levelling bowl of an OBS system to the datum
position. The parameter which should be one of 1BOWL (for

single-bowl units), Z, N/S or E/W.

adc­command module obs­unlock

Unlock each of the masses for deployment of an OBS system.
This sub-command takes no options.

adc­command module obs­lock

Lock each of the masses for deployment of an OBS system.
This sub-command takes no options.

November 2010