The second section of this web page deals with calibration. Note that
all entered values must be integers so, for example, if you wish to
calibrate with a 0.5Hz sine-wave, this should be entered as a 2 second
period. Please refer to the relevant manuals for your digitiser and
sensor for more details of these options.
The screen looks like this (shown in sections):
The Amplitude value refers to an arbitrary full-scale output from the
calibration circuitry. It can be left at 100% unless clipping is
observed, in which case it should be reduced until an undistorted
output is observed.
The square wave (step) and random calibration dialogues are similar: