CMG-EAM (Platinum Firmware)

Perform a sine-wave calibration according to the specified
arguments, which are given as a space-separated list of

key=value pairs:

component=[ ALL | Z | N/S | E/W ] - specify
the component(s) to be calibrated

duration=m - specify the duration, in minutes, where

m is an integer value. For accurate results, the
duration should be significantly longer than the

response time of the instrument.

amplitude=a - specify the amplitude, where a is the
integer percentage of the full output of the cailbration

signal generator. This can normally be given as 100
but should be reduced if clipping is noticed.

freq=n - specify the frequency (in integer Hertz) or

period (in integer seconds) of the calibration signal,
according to the setting of...

freq_or_period=[ Hz | sec ] - specifies whether

the value associated with freq is interpreted as Hertz or
seconds.

adc­command module calib­step arguments

Perform a step calibration according to the specified
arguments, which are given as a space-separated list of

key=value pairs:

component=[ ALL | Z | N/S | E/W ] - specify

the component(s) to be calibrated

duration=m - specify the duration, in minutes, where
m is an integer value. The calibration signal will first

go negative for this duration, then back to zero for the
same duration, then go positive for this duration, then

return to zero again for the same duration. For
accurate results, the duration should be significantly

longer than the response time of the instrument.

amplitude=a - specify the amplitude, where a is the
integer percentage of the full output of the cailbration

signal generator. This can normally be given as 100
but should be reduced if clipping is noticed.