Cable specification, Ups j56 – MTS FlexTest Models 200 User Manual
Page 82

MTS FlexTest® Models 40/60/100/200 Controller Hardware
Model 494.42 Single-Station System Board
FlexTest Controller Configurations
82
J56 UPS Input Connections for the Model 494.42 System Board
Connector J56 UPS provides dedicated inputs for an uninterruptible power
supply (UPS).
Cable specification
To maintain EMC compliance, the J56 UPS cable must comply with the
following specifications:
Connector–9-pin contact type D male EMI connector
Back shell–EMI metallized plastic
Cable–AWG and number of conductors as required. Braided shield with shield
connected to the metallized backshell at the chassis.
494.42 System Board
J56
UPS
J56
Reserved
AC Fail
UPS
Open=Logic1=Active
UPS Battery Low
Reserved
Reserved
Reserved
Reserved
UPS AC Fail
Battery Low
Power
100-240 VAC
50-60 Hz, 1-2 A
DA Output