beautypg.com

MTS FlexTest Models 200 User Manual

Page 153

background image

Digital Universal Conditioner Mezzanine Cards

MTS FlexTest® Models 40/60/100/200 Controller Hardware

Mezzanine Cards

153

3. Set up the mezzanine card(s).

A. (Optional) For transducers that require a ground reference, set the

card grounding DIP switch to the ON position.

B. (Optional) For quarter-bridge transducers, set the bridge-

completion DIP switch on the card to the ON position and install a
bridge completion resistor on the I/O carrier board.

C. (Optional) Install shunt calibration resistor plug assemblies on the

front panel of the I/O carrier board.

Note

If you use MTS TEDS modules or MTS transducers with integrated
shunt-calibration resistors, you must insert a jumper plug (MTS part
number 100-188-097) into the socket for each transducer input
where you will use the integrated shunt-calibration resistor.

4. Install the mezzanine cards on the I/O carrier board.

5. Repeat the preceding steps to set up and install any other cards.

6. Install the I/O carrier board(s).

A. Plug the board into the VME card cage.

B. Connect the transducer cables to the I/O carrier board.

7. Apply power to the system and use hardware-mapping and control

software to configure each conditioner.

Important

If you use a 4-wire cable for a DUC that is configured for an 8-
wire device, the excitation voltage will not be correct.

8. Allow the hardware to warm up for an hour, and then perform

calibration and tuning procedures.