MTS FlexTest Models 200 User Manual
Page 193
![background image](/manuals/567928/193/background.png)
Model 494.47 Dual UART/Encoder Interface Card
MTS FlexTest® Models 40/60/100/200 Controller Hardware
Mezzanine Cards
193
Pulse-Width-Modulated (PWM) Connections for the Model 494.47 Card
I/O Carrier Board
Front-Panel
RJ-50 Connectors
JXA and JXB
- Power
- Data
+ Data
+Power
4
5
6
7
Temposonics G
Transducer
(9 to 28-volt power)
JXA
- Power
- Data
+ Data
+Power
4
5
6
7
Temposonics G
Transducer
(9 to 28-volt power)
JXB
Mezzanine
Card
494.47
Dual
UART/
Encoder
Conditioner
Cable Grounding
1 The cable shield connects
to the metal shielding on
the RJ-50 cable plug.
2 The cable plug shielding
connects to the I/O carrier
board body.
3 The I/O carrier board
connects to earth ground
through the chassis.
Temposonics G (PWM) Connections (9- to 28-volt power option)
JXA
+ Data
- Data
5
6
DC Power +
-
Temposonics G
Transducer
(24-volt Power)
I/O Carrier Board
Front-Panel
RJ-50 Connectors
JXA and JXB
JXB
Mezzanine
Card
494.47
Dual UART/
Encoder
Conditioner
+ Data
- Data
5
6
DC Power +
-
Temposonics G
Transducer
(24-volt Power)
Cable Grounding
1 The cable shield connects
to the metal shielding on
the RJ-50 cable plug.
2 The cable plug shielding
connects to the I/O carrier
board body.
3 The I/O carrier board
connects to earth ground
through the chassis.
4 The negative (-) side of
the DC Power must be
connected to signal
ground on the controller
chassis.
Temposonics G (PWM) Connections (standard 24-volt power option)