MTS FlexTest Models 200 User Manual
Page 69
Model 494.42 Single-Station System Board
MTS FlexTest® Models 40/60/100/200 Controller Hardware
FlexTest Controller Configurations
69
HSM Control*
Off/Low/High
Control
Low Output
High Output
Proportional
Control
Signal Output
Solenoid
Impedance
Ramp Time
(0 to full scale)
Connector J28 (D-9S)
+24 V DC, 1.0 A maximum
+24 V DC, 1.0 A maximum
0–0.78 A
20–
25 Ω
2.1 s or 4.2 s (software selectable)
* The type of HSM control (off/low/high or
proportional) is software configurable.
Load Frame
E-Stop
Crosshead Unlock
Output
Program Interlock
Input
Connector J29 (D15S)
24 V DC at 1 A
Switch Contact Closed = no program interlock
Switch Contact Open = program interlock
Interlock Output
Relay
Voltage
Current
Connector J43 (D9S)
30 V DC/AC maximum
1 A maximum
Normally Open Relay Contacts: Open = Interlock
Normally Closed Relay Contacts: Closed = Interlock
Model 494.42 Specifications (part 2 of 4)
P
ARAMETER
S
PECIFICATION