beautypg.com

MTS FlexTest Models 200 User Manual

Page 266

background image

MTS FlexTest® Models 40/60/100/200 Controller Hardware

Model 494.74 Two-Station HSM Interface Board

Digital I/O and Transition Boards

266

Cable specification

To maintain EMC compliance, J3 Interlock cables must comply with the
following specifications:

Connector–15-contact, type D, male EMI connector.

Backshell–EMI metallized plastic or metal.

Cable–AWG and number of conductors as required. Braided shield with
shield connected to the metallized backshell at the chassis.

Jumper plug required

If connector J3 is not used, you must install a jumper plug to maintain the
integrity of the interlocks. Use jumper plug part number 100-180-351 or
jumper pins: 1 and 4; 2 and 10; 9 and 12.