Digital Universal Conditioner Mezzanine Cards
Mezzanine Cards
146
DC Excitation
1–20 V DC (software configurable).
Important
If you use a 4-wire cable for a DUC
that is configured for an 8-wire device,
the excitation voltage will not be
correct.
Accuracy (1V to 20V): 0.10% of setting + 0.0001
VDC
Loading Effects (100mA max.): 0.02% of setting
‡
Anti Alias Filter
5 pole Butterworth
3 dB = 25kHz
A/D Resolution
16 bit
Excitation
Stability
AC: 30 ppm/
°
C typical
DC: 25 ppm/
°
C typical
AC Excitation
Excitation voltage: 0.5 Vpeak to 10 Vpeak
(software configurable)
Excitation frequencies: 10 KHz, 5 KHz, 2.5 KHz,
2 KHz, 1 KHz (software configurable)
Distortion (THD):
≤
1% typical
Accuracy (2.5V to 10Vpk): 0.30% of setting
§
Loading Effects (100mA max.): 0.15% of setting
#
Excitation Drive
Current
100 mA maximum (AC/DC)
DUC Specifications (part 2 of 3)
P
ARAMETER
S
PECIFICATION