MTS FlexTest® Models 40/60/100/200 Controller Hardware

Digital Universal Conditioner Mezzanine Cards

Mezzanine Cards

146

DC Excitation

1–20 V DC (software configurable).

Important

If you use a 4-wire cable for a DUC
that is configured for an 8-wire device,
the excitation voltage will not be
correct.

Accuracy (1V to 20V): 0.10% of setting + 0.0001
VDC

Loading Effects (100mA max.): 0.02% of setting

Anti Alias Filter

5 pole Butterworth
3 dB = 25kHz

A/D Resolution

16 bit

Excitation
Stability

AC: 30 ppm/

°

C typical

DC: 25 ppm/

°

C typical

AC Excitation

Excitation voltage: 0.5 Vpeak to 10 Vpeak
(software configurable)

Excitation frequencies: 10 KHz, 5 KHz, 2.5 KHz,
2 KHz, 1 KHz (software configurable)

Distortion (THD):

1% typical

Accuracy (2.5V to 10Vpk): 0.30% of setting

§

Loading Effects (100mA max.): 0.15% of setting

#

Excitation Drive
Current

100 mA maximum (AC/DC)

DUC Specifications (part 2 of 3)

P

ARAMETER

S