Model 494.16 vd/duc card settings – MTS FlexTest Models 200 User Manual
Page 164

MTS FlexTest® Models 40/60/100/200 Controller Hardware
Model 494.16 VD/DUC Mezzanine Card
Mezzanine Cards
164
Model 494.16 VD/DUC Card Settings
Software configuration
Most configuration and setup of the valve driver (VD) and digital universal
conditioner (DUC) is done through software.
Note
For detailed information on software configuration, calibration, and
tuning, see the control software and tuning/calibration manuals.
Hardware settings
The conditioner portion of this card has a two-position switch (SW1) that is
used to set up the following DC conditioner functions:
•
Switch 1A connects a bridge-completion resistor to -FBR to complete a
quarter-bridge circuit.
•
Switch 1B provides a ground reference for an external transducer by
connecting -FBR to ground.
1 2
Switch Settings
A = Bridge Completion Switch
The ON position connects a
user-installed bridge completion
resistor to complete a quarter-
bridge circuit.
B = Grounding Switch
The ON position connects -FBR
line to ground. (See the bridge
completion drawings for details.)
SW1