MTS FlexTest Models 200 User Manual

Page 71

background image

Model 494.42 Single-Station System Board

MTS FlexTest® Models 40/60/100/200 Controller Hardware

FlexTest Controller Configurations

71

Digital Output
Relays

Voltage

Current

Output logic

Aux. Voltage

Connector J55 (D9S)

30 V AC/DC maximum

1 A maximum

Output 1: One normally open (NO) contact
Open = Output is OFF

Output 2: One normally open (NO) contact and one
normally closed (NC) contact
NO Contact: Open = Output is OFF
NC Contact: Closed = Output is OFF

Output 3: One normally open (NO) contact
Open = Output is OFF

24 V DC, 0.75 A

UPS Alarm Inputs

Connector J56 (D9S)

UPS ACFAIL Alarm Input (switch contact):
Open=UPS ACFAIL active (logic 1)
Closed= UPS ACFAIL not active (logic 0)

UPS Low Battery Alarm Input (switch contact):
Open=UPS Low Battery active (logic 1)
Closed= UPS Low Battery not active (logic 0)

D/A Outputs

Resolution

Output Type

Output Voltage

Output Current

Connector "D/A Output" (D9P)

16 bit

Single ended

+/- 10 V full scale

5.0 mA maximum

Note

This connector is located on the rear panel
of the Model 494.04 Chassis.

Model 494.42 Specifications (part 4 of 4)

P

ARAMETER

S

PECIFICATION