Rack-Mounting Requirements–Series 494 Chassis
To install a Series 494 Chassis in most 19-inch consoles you will need a console
mounting kit.
Grounding Requirements–Series 494 Chassis
The controller will not function correctly if the chassis is not grounded as shown.
Make sure that your power source is also properly grounded.
The chassis includes two grounds: a chassis ground and a signal common. During
manufacturing, the two grounding lugs are connected together with an external
shorting bar.
M
OUNTING
K
IT
P
ART
N
UMBER
494.04 Chassis Console Mounting Kit
100-152-784
494.06 Chassis Console Mounting Kit
100-174-282
494.10 Chassis Console Mounting Kit
100-183-825
494.20 Chassis Console Mounting Kit
(Vertical Console)
56-819-501
494.20 Chassis Console Mounting Kit
(Half-Height Console)
56-781-303
Shorting Bar
Signal
Common
Chassis
Ground