beautypg.com

MTS FlexTest Models 200 User Manual

Page 54

background image

MTS FlexTest® Models 40/60/100/200 Controller Hardware

Model 494.41 Single-Station System Board

FlexTest Controller Configurations

54

Program Interlock
Input

Input Logic

Connector J29 (D15S)

Switch Contact Closed = no program interlock
Switch Contact Open = program interlock

HPU Outputs
(Start/Low/High)

HPU Output

HPU ON Input

Connector J25 (D15P)

Output Voltage: 24 V DC at 10 mA

Trip Voltage: 0.9–5.5 V DC

Maximum Input Voltage: +26 V DC

HPU Interlock
Inputs (Low Level
and Overtemp)

Trip-point Voltage

Connector J25 (D15P)

18–23 V DC
Maximum Input Voltage: +26 V DC

Auxiliary Power
Outputs

+5 V DC

+15 V DC

-15 V DC

+24 V DC

Connector J49 (D9S)

0.75 A at 40ºC (104ºF)

0.75 A at 40ºC (104ºF)

0.75 A at 40ºC (104ºF)

0.75 A at 40ºC (104ºF)

Model 494.41 Specifications (part 2 of 3)

P

ARAMETER

S

PECIFICATION